Abstract

Mercury in air was determined after collection in potassium permanganate or sodium carbonate solution. The mercury concentration in these solutions was determined in a computerized flow potentiometric stripping analyzer with a 10-μm gold fibre working electrode, a glassy carbon reference electrode and a platinum counter electrode. After sample electrolysis for 1–10 min, stripping was done in a 1 mg l −1 gold(III) solution in 0.01 M nitric acid/0.01 M sodium nitrate with a constant stripping current of 0.50 μA. Results obtained for flue gas samples were in good agreement with results from cold-vapor atomic absorption spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.