Abstract

A new method for the determination of mercury in air has been developed. It combines the adsorption of mercury on Hopcalite (a material approved for this purpose by the National Institute of Health of the United States) and its quantification by neutron activation. The concentrations of mercury in office air in Montreal, Canada, were determined by instrumental semiabsolute neutron activation analysis. They varied from 39 to 48 ng/m 3 . The results were compared with the concentrations of mercury in office air determined simultaneously at the same place by cold vapour atomic fluorescence spectrophotometry. A close correlation between the results of the 2 methods was obtained. The detection limit of our method is about 14 ng/m 3 .

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.