Abstract

Abstract A method has been developed for the determination of low concentrations of mercury in air (nanograms/m3), i.e. in the range of the believed natural levels of mercury in the atmosphere (20 ng/m3). Mercury vapour has been collected from up to 2001 of air in glass tubes containing thin films of gold on sieved ceramic powder. In the laboratory theabsorbed mercury was then released into a quartz-window cell by heating the tube in an oven at 500°C. In this paper it is demonstrated that, by using extremely thin films of precipitated gold, quantitative recovery is obtained and memory effects, which result from the use of thicker films, are avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.