Abstract

Systematic experiments were conducted to evaluate and compare the analytical figures of merit of two reducing agents (SnCl2 and NaBH4) in a continuous flow cold vapor atomic fluorescence mercury analyzer. It was found that sodium tetrahydroborate can efficiently reduce Hg2+ in various environmental samples at a concentration as low as 10 microM (ca. 3.8 x 10(-5)% w/v). Most commonly encountered transition metals (Fe2+, Fe3+, Zn2+, Cu2+, Ni2+, Pb2+ and Cr3+) did not interfere with total Hg determination. No interference from hydride-forming elements (Se4+, Sb3+ and As3+) was observed. Interference caused by Mn2+ and Ag+ could be readily removed by dilution and by using appropriate modification of the reaction matrix. A higher concentration of NaBH4 (0.1 M) is stable for I month when stored in the NaOH matrix (0.2 M) and at low temperature (4 degrees C). A working solution of NaBH4 can be freshly prepared by dilution. With NaBH4, the whole continuous flow system is kept clean much more easily as no precipitate is formed, which in turn considerably reduces memory effects, simplifies analytical operation and reduces the chemical cost six-fold.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.