Abstract
The uncertainty of measurement for well-known neurotransmitters like serotonin, noradrenalin and gamma-aminobutyric acid (GABA), glutamate and melatonin hormone with high performance liquid chromatography−fluorescence detection was calculated after the method validation. Two methods were developed for the determination of the neurotransmitters. A derivatization step was performed for the determination of GABA and glutamate. Sensitivity, method detection limit, limit of quantification, linearity, recovery, interday and intraday precision values were calculated. Low detection limit values were obtained especially for the determination of GABA and glutamate. Then, bottom-up approach was used to calculate measurement uncertainty. The critical stages of the method were evaluated. The major sources of the uncertainty budget were calibration curves, stock solution and recovery. The calculated percentage relative uncertainty values for the compounds changed between 10.1 and 16.7.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.