Abstract

A selective ultraperformance liquid chromatographic (UPLC) method for the determination of magnesium valproate and its process related impurities has been developed. The method includes reversed-phase Acquity BEH C18 column with 100 mm × 2.1 mm i.d. and 1.7 µ particle size. The mobile phase consists of acetonitrile and 5 mM ammonium dihydrogen orthophosphate with pH = 3.0 at 45 : 55 isocratic elution. The flow rate was set at 0.3 mL/min and UV detection was performed at 215 nm. A system suitability test (SST) was developed to govern the quality of the separation. The developed method has been validated further with respect to linearity, accuracy, precision, selectivity, LOD, LOQ, and robustness. Different batches of samples were examined using this method; the method proved to be successful when applied to analyze a marketed magnesium valproate formulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.