Abstract

A high-performance liquid chromatographic method for the determination of lomefloxacin in human plasma has been developed and validated. A solid-phase extraction procedure was used to isolate lomefloxacin from the biological matrix prior to the quantitative analysis. The compound was separated on a Vydac anion-exchange column using acetonitrile—phosphate buffer (pH 7.0) as the mobile phase and quantified by measuring its UV absorbance at 280 nm. The lower limit of detection for the analyte was 0.05 μg ml −1. Enoxacin was used as the internal standard. The calibration graph of the method was linear from 0.1 to 10 μg ml −1 of lomefloxacin in human plasma. This procedure is suitable for pharmacological and pharmacokinetic studies of lomefloxacin.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.