Abstract

An analytical methodology for nifedipine quantitation in plasma by on-line solid-phase extraction (SPE) and high-performance liquid chromatography (HPLC) is described. The SPE cartridges contain C 2 and the analytes nifedipine and nitrendipine (internal standard) are separated on a C 18 column with a mobile phase consisting of acetonitrile–13 m M phosphate buffer pH 7 (65:35, v/v) followed by UV detection at 338 nm. Validation of the method demonstrated good recoveries (>90%), sensitivity (limit of quantification, 2 ng/ml), based on a 500 μl sample volume, accuracy and precision (<5.5% in concentrations greater than the limit of quantitation). This methodology has been used for bioequivalence studies.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call