Abstract
The applicability of hollow-fibre liquid-phase microextraction for extracting 2-chlorovinyldichloroarsine (lewisite 1), bis(2-chlorovinyl)chloroarsine (lewisite 2), tris(2-chlorovinyl)arsine (lewisite 3) and arsenic trichloride from aqueous samples is reported. Parameters affecting the extraction efficiency of these chemicals were optimised. These parameters included the type of derivatising agent, extraction solvent, derivatisation method, pH, ionic strength, stirring speed and extraction time. A linear range between 0.002 and 0.2 μg/mL was established for the lewisites with good square regression coefficients (0.9955-0.9992). Good reproducibility with relative standard deviations (RSDs) from 8 to 10% was achieved. The limit of detection was 0.002 μg/mL for the lewisites and 0.005 μg/mL for arsenic trichloride (3:1 signal-to-noise ratio). The extraction method was validated with a proficiency test sample issued by the Organisation for the Prohibition of Chemical Weapons (OPCW). The rapidity and precision of the new method should help deter against the employment of lewisite as a chemical warfare agent: its use could be confirmed easily from analysis of aqueous samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.