Abstract

Abstract Nineteen laboratories, using 6 different models of atomic absorption spectrophotometers and 4 different types of polarographs, participated in this collaborative study. The average lead recoveries from 6 paired samples at 1–2, 5–6, and 10–11 ppm levels were 97.7% by polarography and 100.7% by atomic absorption. The average standard deviations were 0.32 and 0.41 ppm, respectively, and the average coefficients of variation were 7.9 and 13.1%, respectively. With collaborators reporting on both methods, the results of the overall method average were 4.3 ppm for polarography and 4.4 ppm for atomic absorption. Since there were no significant differences (p > 0.05) found between the method averages, except at one of the unspiked levels, the 2 methods can be used to confirm each other at levels of 1-11 ppm. The polarographic method has been adopted as official first action for the determination of lead in fish.