Abstract

A procedure for the simultaneous determination of lead and cadmium in TiO 2 by differential pulse anodic stripping voltammetry (ASV) has been developed. The key feature of the method is the use of triethanolamine (TEA) to remove titanium interference: TiO 2 undergoes acidic digestion with HF/H 2SO 4 at atmospheric pressure, TEA is added to the HCl solution of the residue and the solution is analysed using a standard ASV instrumentation, equipped with a hanging mercury drop electrode. The calibration curves for both lead and cadmium are linear up to 50 μg l −1 of solution, and the detection limits are 1 μg l −1, corresponding to 1 μg g −1 of TiO 2. Method reliability was tested by comparing the results with those given by electrothermal atomic absorption spectroscopy. The method has been successfully applied for determination of both contaminants in powdered titanium dioxide (raw materials) and in titanium dioxide-containing cosmetics (sunscreen products).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.