Epithermal instrumental neutron activation analysis (EINAA) technique in conjunction with anti-coincidence gamma-ray spectrometry (AC) has been applied for the determination of ppm to ppb levels of iodine in biological materials containing high levels of Al, Br, Cl, K, Mn, and Na. Both conventional EINAA-AC and pseudo-cyclic EINAA-AC (PC-EINAA-AC) methods using a combination of Cd and B filters have been developed using Dalhousie University SLOWPOKE-2 reactor (DUSR) facility. The expanded uncertainties (EU), at about 95% confidence level, for iodine in biological materials by EINAA-AC varied between 6 and 10%. The advantages of the non-destructive PC-EINAA-AC method has been successfully demonstrated by analyzing the NIST Pine Needles (SRM 1575) containing a low amount of iodine in presence of high quantities of Mn and other interfering elements where an iodine content of 92.8 μg kg−1 with an EU of 6.1 μg kg−1 and a detection limit of 40 μg kg−1 has been obtained at the end of fourth cycle.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.