Abstract
A set of unirradiated Low Gain Avalanche Detectors from Hamamatsu Photonics from the HGTD prototype 2 run for the upgrade of experiments at the High Luminosity Large Hadron Collider were used to determine the impact ionization parameters for silicon, particularly in the electric field range of ∼30 V/μm which is of interest for LGADs. The parameters' dependence on temperature was also determined. Their validity on irradiated sensors up to 2.5·1015 cm-2 was also tested and found to be acceptable at low biases but further work should be done to determine the dependence on fluence.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.