Abstract
The single ion activity coefficients of hydrogen and chloride ions in aqueous HCl solutions have been estimated at 25°C at concentrations up to 1 mol-kg−1, using potentiometric measurements with ion-selective electrodes and appropriate calibration procedures. Two methods are described for an internal calibration of the electrodes in the extended Debye–Huckel concentration range. The results are compared to the conventional pH calibration with external buffer solutions. Since the latter calibration method does not account for the liquid junction potential E J which arises at the reference electrode, the resulting activity coefficients are quite different in HCl solutions of higher concentration. These differences between internal and external calibration decrease significantly, when a correction for E J is introduced into the conventional pH calibration. Hence, in solutions of higher ionic strength the accuracy of the conventional pH electrode calibration using buffer solutions is very limited, when exact H+ activities are required. The consistency of the results indicates that the liquid junction potentials in the examined systems calculated by the Henderson/Bates approximation are of reasonable precision.