Abstract
We have determined the glucose concentration of whole blood from mid-infrared spectra without sample preparation or use of chemical reagents. We selected 1119-1022 cm(-1) as the optimal wavelength range for our measurement by making a first-loading vector analysis based on partial least-squares regression. We examined the influence of hemoglobin on samples by using different calibration and prediction sets. The accuracy of glucose prediction depended on the hemoglobin level in the calibration model; the sample set should represent the entire range of hemoglobin concentration. We obtained an accuracy of 5.9% in glucose prediction, and this value is well within a clinically acceptable range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.