Abstract

An immunosensors/sequential injection analysis system is proposed for the assay of thyroid hormones free L-T4 and free L-T3 in blood samples. The working concentration ranges for free L-T4 and free L-T3, are between 2.0 and 720.0 pmol/L and between 2.0 and 340.0 pmol/L, respectively, with detection limits of 1.0 and 1.7 pmol/L, respectively. The selectivity of the immunosensors, as well as the recovery tests done for free L-T4 and free L-T3 (99.90 ± 0.01% (n = 10) and 99.94 ± 0.02% (n = 10)), made the immunosensors/sequential injection analysis system suitable for the direct assay of thyroid hormones in blood samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.