Abstract
A novel flow injection chemiluminescence method is proposed for determination of cholesterol in this paper. The cholesterol oxidase was immobilized onto sol-gel and prepared as an enzymatic reaction column. The determination of cholesterol was performed by quantitative determination of hydrogen peroxide produced from an enzymatic reaction. The luminol-H2O2-metal chelate diperiodatocuprate(III) system ensured that the method was highly sensitive and selective. Free cholesterol was determined over the range 5.0 x 10(-8) mol/L-5.0 x 10(-7) mol/L, with a limit of detection (3sigma) of 1.9 x 10(-8) mol/L. The relative standard deviation (RSD) for 2.5 x 10(-7) mol/L was 2.7% (n = 7). The proposed method offered the advantages of sensitivity, selectivity, simplicity and rapidity for free cholesterol determination, and was successfully applied to the direct determination of free cholesterol in serum.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.