Abstract

The quantity of formate was determined using apparatus comprised of a reactor with immobilized formate dehydrogenase (FDH) in the flow line. NADH formed by an enzymatic reaction was fluorometrically detected. The optimal concentration of NAD + in the carrier was determined. The maximum peak area due to NADH was observed at pH 7.0 when the pH of the carrier consisting of piperazine1,4-bis(2-ethanesulfonic acid) (PIPES) buffer ranged from 6.0 to 8.0. Various buffer types were also examined as carrier media at pH 7.0 and PIPES buffer showed the maximum peak area. When the carrier composed of PIPES buffer (0.1M, pH 7.0) was used, the calibration curve for formate was linear in the range of 0.5‐50µ M( r = 1.000). Relative standard deviations of the peak area at 1µ Ma nd 10µM were 2.6% (n = 7) and 1.6% (n = 7), respectively. This method was applied to the analysis of formate in foodstuffs, and formate content determined by this method agreed with that determined by a commercially available test-kit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.