Abstract

A rapid and sensitive liquid chromatography-tandem mass spectrometry method has been developed and validated for the determination of febuxostat in human plasma. The liquid-liquid extraction technique was used for the extraction of febuxostat from human plasma using trandolapril as the internal standard (IS). Chromatography was performed on a ultra-performance liquid chromatography (UPLC) BEH C18, 50 mm X 2.1 mm, 1.7 µm particle size column, with the mobile phase consisting of 0.1% formic acid and acetonitrile (in a 25:75 ratio), followed by detection using mass spectrometry. The method involves a simple reversed isocratic chromatography condition and mass spectrometry detection, which enables detection at sub-microgram levels. The method was validated and the lower limit of quantification for febuxostat was found to be 0.075 µg/ml. The mean recovery for febuxostat ranged from 100.9 to 106.5%. This method increased the sensitivity and selectivity; resulting in high-throughput analysis of febuxostat using commercially available IS for pharmacokinetic, bioavailability, and bioequivalence studies, with a chromatographic run time of 1.5 min only.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call