Abstract

A flow injection method (FI) has been developed for the determination of ethylenediaminetetraacetic acid (EDTA) by flame atomic absorption spectrometry (FAAS). In the first step a copper(II) regenerant solution was injected on to an on-line column packed with Chelex-100 resin. The excess of copper ion, not retained on the column, was washed by an ammonia solution carrier. In the second step an EDTA sample was injected. The analyte displaced an equivalent amount of copper(II) from the chelating column. Eluted copper was determined by an FAA spectrometric detector positioned downstream. The signal obtained was proportional to the concentration of EDTA present in the injected sample. Using this FI-FAAS conversion method the detection limit was 0.1 µg ml–1. The precision of the technique was better than a relative standard deviation of 1.5% at 0.50 µg ml–1 levels, with a throughput of 45 samples h–1. The effect of sample volume and of ionic and organic ligand interferents on the FI-FAAS signals was studied. Only glycylglycine was found to interfere significantly.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.