The sensitivity and selectivity of gas chromatography for analysing several halogenated ester derivatives of β-hydroxymyristic acid were studied using both selected-ion monitoring detection with negative-ion chemical-ionization mass spectrometry and electron-capture detection. Six different derivatization methods were compared with respect to yield, chemical stability and formation of by-products. Procedures for removal of excess reagents using disposable silica columns and thin-layer chromatography were elaborated. The 3-O-pentafluorobenzoyl-methyl ester was the preferred derivative since it provided high sensitivity and had the molecular ion as the base peak in the mass spectrum. The detection limit was 0.5 pg with electron-capture detection and 0.3 pg with the mass spectrometric system. Using β-hydroxymyristic acid as a chemical marker it was possible to detect Escherichia coli endotoxin in aqueous solution at a level of 1 ng/ml.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.