Abstract
Cold vapor cavity ringdown spectroscopy has been successfully applied to the detection of elemental mercury. Using an absorption cell 0.18 m in length, detection limits of 0.027 and 0.12 ng were obtained using peak area and peak height measurements, respectively. For the peak area measurement, this corresponds to a gas phase concentration of less than 25 ng m−3. For comparison, using a similar absorption cell, standard AAS yielded a Hg detection limit (peak height) of 9 ng, (gas phase concentration of ≡ 8.3 μg m−3).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.