Abstract

The effective energy of a polychromatic beam for a Computed Tomography (CT) scanner can be measured directly only with difficulty. However, a linear relationship exists between the measured CT numbers and corresponding attenuation coefficients of known materials at the effective energy of the x-ray beam. The effective energy can then be determined by searching all energies for the best linear correlation between the CT numbers and the attenuation coefficients. This can be performed by two methods: graphically, by means of choosing visually the straightest of the fitted lines or, mathematically, by maximizing the correlation coefficient. The energy corresponding to the optimal fit is therefore selected as the effective energy. The latter method was implemented by computer and demonstrated by scanning the AAPM phantom, which contained known materials, and determining the effective energies and the relationship between the linear attenuation coefficients and CT numbers for three commercial units.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.