Abstract

PurposeIn photon radiation therapy, computed tomography (CT) numbers are converted into values for mass density (MD) or relative electron density to water (RED). CT‐MD or CT‐RED calibration tables are relevant for human body dose calculation in an inhomogeneous medium. CT‐MD or CT‐RED calibration tables are influenced by patient imaging (CT scanner manufacturer, scanning parameters, and patient size), the calibration process (tissue‐equivalent phantom manufacturer, and selection of tissue‐equivalent material), differences between tissue‐equivalent materials and standard tissues, and the dose calculation algorithm applied; however, a CT number calibration audit has not been established. The purposes of this study were to develop a postal audit phantom, and to establish a CT number calibration audit process.MethodsA conventional stoichiometric calibration conducts a least square fit of the relationships between the MD, material weight, and measured CT number, using two parameters. In this study, a new stoichiometric CT number calibration scheme has been empirically established, using three parameters to harmonize the calculated CT number with the measured CT number for air and lung tissue. In addition, the suitable material set and the minimal number of materials required for stoichiometric CT number calibration were determined. The MDs and elemental weights from the International Commission on Radiological Protection Publication 110 were used as standard tissue data, to generate the CT‐MD and CT‐RED calibration tables. A small‐sized, CT number calibration phantom was developed for a postal audit, and stoichiometric CT number calibration with the phantom was compared to the CT number calibration tables registered in the radiotherapy treatment planning systems (RTPSs) associated with five radiotherapy institutions.ResultsWhen a least square fit was performed for the stoichiometric CT number calibration with the three parameters, the calculated CT number showed better agreement with the measured CT number. We established stoichiometric CT number calibration using only two materials because the accuracy of the process was determined not by the number of used materials but by the number of elements contained. The stoichiometric CT number calibration was comparable to the tissue‐substitute calibration, with a dose difference less than 1%. An outline of the CT number calibration audit was demonstrated through a multi‐institutional study.ConclusionsWe established a new stoichiometric CT number calibration method for validating the CT number calibration tables registered in RTPSs. We also developed a CT number calibration phantom for a postal audit, which was verified by the performances of multiple CT scanners located at several institutions. The new stoichiometric CT number calibration has the advantages of being performed using only two materials, and decreasing the difference between the calculated and measured CT numbers for air and lung tissue. In the future, a postal CT number calibration audit might be achievable using a smaller phantom.

Highlights

  • Computed tomography (CT) images — required for contouring a target and calculating dose distribution in a patient’s body — are imported into radiotherapy treatment planning systems (RTPSs)

  • CT numbers are converted to mass density (MD), or relative electron density to water (RED), according to the applicable RTPS or the dose calculation algorithm, and the dose distribution is calculated for the human body in an inhomogeneous medium

  • Dose calculation in an inhomogeneous medium is influenced by four factors: patient imaging, the calibration process, the difference between tissue-equivalent material and standard tissues, and the dose calculation algorithm applied

Read more

Summary

Introduction

Computed tomography (CT) images — required for contouring a target and calculating dose distribution in a patient’s body — are imported into radiotherapy treatment planning systems (RTPSs). Dose calculation in an inhomogeneous medium is influenced by four factors: patient imaging (itself influenced by CT scanner manufacturer, scanning parameters, and patient size), the calibration process (influenced by tissue-equivalent phantom manufacturer and selection of tissue-equivalent materials), the difference between tissue-equivalent material and standard tissues, and the dose calculation algorithm applied. To validate dose calculation accuracy in the inhomogeneous medium, a comparison between the calculation and measurements, using lung or bone equivalent phantoms, is usually conducted.[1,2,3,4,5] Final dose calculation results in the inhomogeneous medium vary according to the four factors above; practitioners such as radiation therapists or medical physicists can only adjust two of these, patient imaging and the calibration process. It is valid to review patient imaging and the calibration processes using a third party; such patient imaging and calibration process reviews have not been performed

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call