Abstract

The conditions for the determination of copper and manganese in seawater by high-resolution continuum-source graphite furnace atomic absorption spectrometry (HR-CS GFAAS) were optimized. The analytical properties were compared with those of low-resolution line source GFAAS (LR-LS GFAAS). The metrological characteristics were estimated considering the dilution of the sample and optimization of the temperature program of the electrothermal atomizer. Using HR-CS GFAAS, the elements were determined in undiluted samples, while LR-LS GFAAS allows the determination only at double dilution. The lower limits of determined Cu concentrations almost coincided for both versions of GFAAS. For manganese, the limits of detection and quantification were significantly lower for HR-CS GFAAS. Calibration curves were plotted using standard solutions. The proposed scheme was tested on water samples from the Sea of Okhotsk by standard addition. The detection limits (LODs) were 0.6 µg L−1 for copper and 0.07 µg L−1 for manganese by HR-CS GFAAS and 0.5 µg L−1 for copper and 0.13 µg L−1 for manganese by LR-LS GFAAS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.