Abstract

An optimized method for the determination of flecainide in serum is presented. Extraction using a solid-phase C 18 column and chromatography on a stabilized fluorocarbon-bonded silica gel column effectively separate flecainide from an internal standard (a positional isomer of flecainide). The HPLC apparatus and conditions were as follows: analytical column, Fluofix 120N; sample solvent, 20 μl; column temperature, 40°C; detector, Shimadzu RF-5000 fluorescence spectrophotometer (excitation wavelength=300 nm, emission wavelength=370 nm); mobile phase, 0.06% phosphoric acid containing 0.1% tetra- n-butyl ammonium bromide–acetonitrile (75:25, v/v); flow-rate, 1.0 ml/min. The standard curves for flecainide were linear in the concentration range examined (10–2000 ng/ml). The regression equation was y=0.08+0.0078 x ( r=0.9998). The minimum detectable amount of flecainide was approximately 5 ng/ml. In the within-day study, the precision coefficients of variation were 2.66, 2.18, 2.54, 2.72, 2.88, 2.24, and 3.29% for the 10, 50, 100, 200, 500, 1000, and 1500 ng/ml standards, respectively. The absolute recovery rates of flecainide at each concentrations were 94–100%. The method described provides analytical sensitivity, specificity and reproducibility suitable for both biomedical research and therapeutic drug monitoring.