Abstract
Abstract A method is proposed for the solvent extraction of cobalt using 1,5-bis(di-2-pyridylmethylene) thiocarbonohydrazide (DPTH) as extractant. The optimum extraction conditions were evaluated from a critical study of the effect of pH. concentration of extractant, shaking time and ionic strength. The maximum volume ratio of the aqueous to organic phase was 30:1 for a single-stage extraction of 99-100 % of the metal ion. The detection limit is 0.9 ng/ml cobalt, and the calibration is linear from 0.9 to 1000 ng/ml. The effect of interferences was studied and no interferences from the elements commonly found in biological materials were observed. The extraction method was applied to the determination of cobalt in some biological materials using inductively coupled plasma atomic emission spectrometry (ICP-AES).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.