Abstract

This study was aimed at examining two flow injection-spectrophotometric systems (normal and reverse) for the determination of clonazepam (CLO) at the microgram level in pure and pharmaceutical dosage forms. The estimation of CLO has been developed by conjugating a normal (or reverse) flow injection analysis (nFIA or rFIA) and spectrophotometric detection with phloroglucinol as a coupling reagent. Beer’s law was obeyed over a range of 50–400 and 30–400 μg/mL. The limits of detection were 11 and 8 μg/mL and the sampling rates were 51 and 28 samples per hour for nFIA and rFIA respectively. Both systems were successfully applied for the determination of CLO in its commercially available dosage forms. A comparison between the proposed flow systems was also done. These simple and high throughput methods could be utilized for pharmaceutical analysis of CLO.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.