Abstract

A flow injection irreversible biamperometric method for the determination of chlorogenic acid is described. The proposed method is based on the electrochemical oxidation of chlorogenic acid at pretreated platinum electrode and the reduction of permanganate at another electrode to form an irreversible biamperometric detection system. Under the external potential difference (Δ E) of 0 V, in the 0.05 mol/L sulfuric acid, chlorogenic acid can be determined over the range 0.8–120 mg/L with a sample measurement frequency of 80 samples/h. The detection limit is 0.18 mg/L. The proposed method exhibits the satisfactory reproducibility with a relative standard derivation (R.S.D.) of 2.21% for 19 successive determinations of 40 mg/L.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.