Abstract

An assay procedure is described to quantitate low picogram concentrations of catecholamines in human plasma using reversed phase ion-pair HPLC with electrochemical detection. Optimization of chromatographic conditions allows use of samples as small as 0.25 mL. The overall recovery of 78% offers high precision and accuracy at low concentrations. The lower limit of detection is 0.87 pg/mL for epinephrine, 3.5 pg/mL for norepinephrine and 8.3 pg/mL for dopamine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.