Abstract

A potentiometric flow injection‐type biosensor developed in our laboratory was used for the determination of carbamate pesticides. The principle of the biosensor is based on the inhibition of enzyme by pesticides, which is dependent on the concentration of the pesticides present. The sensor system consisted of a reactor with acetylcholinesterase immobilized on controlled pore glass and a detector with tubular H+‐selective membrane electrode. The sensor protocol established was applied to the analysis of nine carbamate pesticides. All the pesticides at 10−6 M inhibited the sensor enzyme, demonstrating the utility of the developed method. Treatment of the inhibited enzyme with pyridine‐2‐aldoxime restored nearly all of the enzyme activity allowing repeated use of the sensor. Calibration curves for carbaryl and carbofuran over the concentration range 10−11–10−4 M were obtained. The lower detection limits were 10−10 M and 10−11 M, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.