Abstract

This paper describes the application of the method of post-column derivatization with triiodide and UV detection at 352 nm for the determination of bromate in drinking water, mineral water and swimming pool water samples. Optimized analytical conditions were further validated in terms of accuracy, precision, linearity, limit of detection and limit of quantification. The method detection limit was at the level of 0.4 μg/L, and the spiked recovery for bromate was in the range of 92% – 104%. The method did not need a special sample treatment and was successfully applied to the analysis of bromate at the required value that is below 2.5 μg/L.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.