Abstract

A rapid, simple, precise and accurate high performance liquid chromatographic (HPLC) analytical method was developed and validated for the determination of the active substances (a.s.) azoxystrobin, topramezone, acetamiprid, fluometuron and folpet in their respective commercially available formulations. The method was used for the analysis of samples under the frame of the national quality control program of plant protection products in the Greek market. Chromatographic separation of the active substances from additives and co-formulants is achieved using isocratic elution with acetonitrile and 0.1% phosphoric acid solution (60:40 v/v) at a flow rate of 0.4 mL min−1 on a C18 monolithic column (Chromolith Performance-RP18e 100 × 4.6 mm) and UV detection at 230 nm. Validation parameters of the method fulfilled acceptability criteria. Recovery of all individual compounds was in the range 97.8–100.9%. Precision expressed as relative standard deviation was lower than the theoretical values of the modified Horwitz equation. Correlation coefficients of linearity of response were better than 0.999. The benefits of the proposed method are significant reduction in analysis time and total cost since all analytes were determined with the same extraction procedures and chromatographic system. Analysis of real samples for all active ingredients confirmed fitness for purpose of the suggested method.