Abstract
A sensitive and accurate liquid chromatographic method for the determination of AR-42 enantiomeric purity has been developed and validated. Baseline separation with a resolution higher than 1.9 was accomplished within 10 min using a CHIRALPAK AD column (250 mm × 4.6 mm; particle size 5 μm) and n-hexane/2-propanol/diethylamine (75:25:0.1 v/v/v) as mobile phase at a flow rate of 1 mL min−1. Eluted analytes were monitored by UV absorption at 260 nm. The effects of mobile phase components, temperature and flow rate on enantiomeric selectivity and resolution of enantiomers were investigated. Calibration curves were plotted within the concentration range between 0.001 and 0.5 mg mL−1 (n = 10), and the recoveries between 98.23 and 101.87% were obtained, with relative standard deviation lower than 1.31%. Limit of detection and limit of quantitation for AR-42 were 0.39 and 1.28 μg mL−1 and for its enantiomer were 0.36 and 1.19 μg mL−1, respectively. It was demonstrated that the developed method was accurate, robust and sensitive for the determination of enantiomeric purity of AR-42, especially for the analysis of bulk samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.