Abstract

A precise and sensitive LC method for the determination of repertaxin enantiomeric purity has been developed and validated. Baseline separation with a resolution higher than 2.0 was accomplished within 20 min using a Chiralpak AD-H column (250 × 4.6 mm; particle size 5 μm) and n-hexane:2-propanol (90:10 v/v) as mobile phase at a flow rate of 1 mL min−1. Eluted analytes were monitored by UV detection at 260 nm. The effects of mobile phase composition, temperature and flow rate on enantiomeric selectivity and on resolution of enantiomers were investigated. Calibration curves were plotted within the concentration range between 0.002 and 1.0 mg mL−1 (n = 3), and relative standard deviation (RSD) of the inter-batch assay and intra-batch assay was less than 1.27 and 1.16 %. LOD and LOQ for repertaxin were 0.65 and 2.19 μg mL−1; those for its enantiomer were 0.70 and 2.34 μg mL−1, respectively. The method was evaluated and validated by analysis of bulk samples of repertaxin of different enantiomeric purity. It was demonstrated that the method was accurate, robust, and sensitive, and enabled practical analysis of real samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.