Abstract

A double column-switching HPLC method with naphthalene-2,3-dialdehyde (NDA) was applied for determination of two plasma antihypertensive peptides, Val-Tyr (VY) and Ile-Val-Tyr (IVY). After a first separation on a Phe-ODS column, double heart-cuts of the retention time corresponding to NDA-VY and NDA-IVY elutions were successfully separated on an analytical ODS column: 60% acetonitrile in 0.1% trifluoroacetic acid containing 5 mM sodium 1-octanesulfonate at 1.0 mL/min. Within-run coefficients of variation were 1.73% and 4.73% for VY and IVY, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.