Abstract

In this study, we developed and validated microanalysis methods for the determination of linear alkylbenzenesulfonate (LAS), sodium lauryl sulfate (SLS), and alpha olefin sulfonate (AOS). The conditions for the analysis of the surfactants using HPLC with FLD, RID, and ELSD detectors were investigated. The methods were validated by determining the linearity, limits of detection (LODs), limits of quantification (LOQs), recovery, precision, and accuracy. LAS analysis by FLD revealed calibration curves that were linear in the range of 10-200 mg/L for an LAS mixture. The calibration curves for C10-C13 had correlation coefficients of 0.995, 0.997, 0.996, and 0.997, respectively. SLS analysis using RID generated a linear calibration curve in the range of 10-300 mg/L. The calibration curve for SLS C12 had a correlation coefficient of 0.9994. AOS analysis using ELSD resulted in a correlation coefficient of 0.9940. For LAS, the LODs and LOQs were 0.09-0.56 and 0.30-1.87 mg/L, respectively. For SLS C12, the LOD and LOQ were 0.07 and 2.33 mg/L, respectively. For AOS C14, the LOD and LOQ were 16.55 and 21.83 mg/L, respectively. The recoveries were 97.17-98.84% for LAS C10-C14, 97.94% for SLS C12, and 96.11% for AOS C14. The established methods provide acceptable precision and accuracy. Our methods could be useful for the detection of anionic surfactants in dishwashing detergents.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call