A method for direct determination of amoxicillin in pharmaceuticals using flow injection analysis (FIA) with UV detection adjusted at 228 nm, along with potentiometer and condoptometric methods are described. The best carrier solvent system consisted of an aqueous solution of methanol (10% v/v) with pH 9 pumped at a flow rate of 1 mL · min−1. The limit of detection (LOD) and quantification (LOQ) for FIA was calculated to be 3.3×10−7 M (S/N=3) and 1.0×10−7 M (S/N=10), respectively. Two pharmaceutical formulations containing amoxicillin were studied. In the analysis of the tablets, the RSD values were found to be 1.36 and 1.21 and 1.18 (for AMOKLAVIN®), 1.10 1.16 and 1.14 (for REMOXIL®) for FIA, potentiometer, and conduct metric methods, respectively. The proposed methods were applied to the pharmaceutical tablets and good results were obtained. These proposed methods are simple, fast, accurate, precise, and cost effective for the routine determination of amoxicillin in pharmaceutical preparations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.