Abstract
Methods based on the blank signal as proposed by IUPAC procedure and on the signal to noise ratio (S/N) as listed in the ISO-11843-1 norm for determination of the limit of detection (LOD) and quantitation (LOQ) in one-dimensional capillary gas chromatography (1D-GC) and comprehensive two-dimensional capillary gas chromatography (CG×GC) are described in detail and compared for both techniques. Flame ionization detection was applied and variables were the data acquisition frequency and, for CG×GC, also the modulation time. It has been stated that LOD and LOQ estimated according to IUPAC might be successfully used for 1D-GC–FID method. Moreover, LOD and LOQ decrease with decrease of data acquisition frequency (DAF). For GC×GC–FID, estimation of LOD by IUPAC gave poor reproducibility of results while for LOQ reproducibility was acceptable (within ±10% rel.). The LOD and LOQ determined by the S/N concept both for 1D-GC–FID and GC×GC–FID methods are ca. three times higher than those values estimated by the standard deviation of the blank. Since the distribution pattern of modulated peaks for any analyte separated by GC×GC is random and cannot be predicted, LOQ and LOD may vary within 30% for 3s modulation time. Concerning sensitivity, 1D-GC–FID at 2Hz and of GC×GC–FID at 50Hz shows a ca. 5 times enhancement of sensitivity in the modulated signal output.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.