Abstract

The analytical performance of laboratories participating in the serum aluminium programme of the Guildford Trace Element Quality Assessment Scheme has been evaluated. The considerable disparity between reported results was not attributable to the nature of the samples distributed, nor to the standard aluminium solutions used in different laboratories. All laboratories used electrothermal atomic absorption spectrometry but sample preparation procedures and temperature programmes varied enormously. While some laboratories performed better than others in the scheme, and the standard of analysis appears to be improving with time, no laboratory showed consistent acceptable performance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.