Abstract

A procedure is described for the determination of aluminium in samples of human origin. It is based on Zeeman-corrected atomic absorption spectrometry, and the method of standard additions is used for the quantitation. Plasma and urine samples were analysed directly after dilution and tissue samples after wet digestion with nitric acid. For plasma samples, the precision was approximately 5%(relative standard deviation) at the 50 µg l–1 level, and the accuracy of the method was evaluated by analysing the NBS reference material bovine serum. For tissue samples, the accuracy and precision were determined from recovery experiments. Examples of aluminium concentrations found in human plasma, urine, bone, liver and brain are given.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.