Optimization of the furnace parameters for electrothermal atomization of mercury leads to a characteristic mass of 20 pg in aqueous solution and 30 pg in chloroform extracts (with Zeeman correction). With a single-step solvent extraction from 100 ml of sample, performed in the sampling vessel, and direct injection of 400 μl of the extract into the furnace, a characteristic concentration of ∼0.8 ng/l. is reached.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.