Abstract

Quality control on second-generation lithotripters has received scant attention in the medical literature. Serial acoustic output measurements were performed on both shock heads of the Siemens lithotripter (Lithostar and Lithostar Plus) in Bristol to identify any variation in the shock wave parameters during the lifetime of the shock head. This testing confirmed a drop in output to less than 50% of the original output on several occasions. The lifetime of the shock head was quite variable, ranging from 37,000 to 727,000 shocks. Serial acoustic output measurements should be mandatory on commercial lithotripters if quality assurance is to be maintained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.