Abstract

A non-uniform response across scanning transmission electron microscope annular detectors has been found experimentally, but is seldom incorporated into simulations. Through case study simulations, we establish the nature and scale of the discrepancies which may arise from failing to account for detector non-uniformity. If standard detectors are used at long camera lengths such that the detector is within or near to the bright field region, we find errors in contrast of the order of 10%, sufficiently small for qualitative work but non-trivial as experiments become more quantitative. In cases where the detector has been characterized in advance, we discuss the detector response normalization and how it may be incorporated in simulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.