Abstract

In the current study, performance of electrochemiluminescence immunoassay (ECLIA) in detection of anti-toxoplasma IgG in human sera was compared with that of enzyme-linked immunosorbent assay (ELISA). Furthermore, performance of an in house Dot-ELISA in detection of anti-toxoplasma IgG was compared with that of ECLIA and ELISA. In total, 219 human sera were tested to detect anti-toxoplasma IgG using Dynex DS2® and Roche Cobas® e411 Automated Analyzers. Discordant results rechecked using immunofluorescence assay (IFA). Then, sera were used in an in house Dot-ELISA to assess toxoplasma-specific IgG. Of the 219 samples, two samples were found undetermined using ECLIA but reactive using ELISA. Using IFA, the two sera were reported unreactive. Furthermore, two samples were found reactive using ECLIA and unreactive using ELISA. These samples were reported reactive using IFA. The overall agreement for the two former methods was 98% (rZ0.98.1; P< 0.001). The intrinsic parameters calculated for in house Dot-ELISA included sensitivity of 79.5, specificity of 78.2, and accuracy of 78.9%, compared to ECLIA and ELISA. Positive and negative predictive values included 82.9 and 74.2%, respectively. A 100% sensitivity was found in in house Dot-ELISA for highly reactive sera in ECLIA and ELISA. ECLIA is appropriate for the first-line serological screening tests and can replace ELISA due to high speed, sensitivity, and specificity, particularly in large laboratories. Dot-ELISA is a rapid, sensitive, specific, cost-effective, user-friendly, and field-portable technique and hence can be used for screening toxoplasmosis, especially in rural fields or less equipped laboratories.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call