A light-addressable potentiometric (silicon) sensor was used in an immunofiltration procedure for the detection of pathogenic bacteria. Yersinia pestis was detected by filtering the cells onto nitrocellulose membranes and then filtering anti-Y. pestis mouse monoclonal antibody and anti-mouse immunoglobulin G-horseradish peroxidase conjugate. For Neisseria meningitidis detection, mouse monoclonal antibody to the major outer membrane protein of this bacterium was coupled directly to horseradish peroxidase. N. meningitidis cell suspensions were filtered onto polycarbonate membranes, and the enzyme conjugate was allowed to react with the filtered bacteria. The presence of both enzyme conjugates was determined potentiometrically with the silicon sensor. The sensitivity of this technique relative to that of an enzyme-linked immunosorbent assay for N. meningitidis was determined. Fewer than 1,000 bacterial cells could be detected with the silicon sensor in a 20-min assay, whereas a 2.5-h enzyme-linked immunosorbent assay with the same antigen and antibody preparations was significantly less sensitive.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.