Abstract
Clotting is a physiological process that prevents blood loss after injury. An imbalance in clotting factors can lead to lethal consequences such as exsanguination or inappropriate thrombosis. Clinical methods to monitor clotting and fibrinolysis typically measure the viscoelasticity of whole blood or optical density of plasma over time. Though these methods provide insights into clotting and fibrinolysis, they require milliliters of blood which can worsen anemia or only provide partial information. To overcome these limitations, a high-frequency photoacoustic (HFPA) imaging system was developed to detect clotting and lysis in blood. Clotting was initiated in vitro in reconstituted blood using thrombin and lysed with urokinase plasminogen activator. Frequency spectra measured using HFPA signals (10–40 MHz) between non-clotted blood and clotted blood differed markedly, allowing tracking of clot initiation and lysis in volumes of blood as low as 25 µL/test. HFPA imaging shows potential as a point-of-care examination of coagulation and fibrinolysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.