Abstract

A liquid chromatography-electrochemistry (LC-EC) method is described for the determination of basal acetylcholine (ACh) in microdialysate from the striatum of freely moving rats. This method is based on the separation of ACh and choline (Ch) by microbore liquid chromatography followed by passage of the effluent through a post-column immobilized enzyme reactor (IMER), containing acetylcholinesterase (AChE) and choline oxidase (ChO), and then the electrochemical detection of the hydrogen peroxide produced. Instead of the conventional platinum electrode used for the anodic detection of hydrogen peroxide, a peroxidase-redox polymer modified glassy carbon electrode operated at + 100 mV vs. Ag/AgCl has been used to detect the reduction of hydrogen peroxide. With this method, a detection limit of 10 fmol (injected) for ACh ( S/ N = 3:1) was obtained and the basal ACh concentration in striatal microdialysate was determined without using esterase inhibitors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.