Abstract

The detection limit is an important figure of merit for evaluating instrumentation and analytical methods. While the detection limit for techniques using linear calibration functions has been studied extensively, this fundamental metric has rarely been discussed for mass spectrometry that bases the calibration on the principle of isotope dilution. We have developed a formulation for the detection limit for isotope dilution mass spectrometry (IDMS) after a thorough analysis of the uncertainty of IDMS measurements. The new formulation describes the IDMS detection limit as a function of the enrichment of the isotopic spike and the linear calibration detection limits measured at the masses for the isotope ratio measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.