Abstract

Rare, heterogeneously composed platinum group element alloy micronuggets (PGNs) occur in primitive meteorites, micrometeorites, and terrestrial impactite deposits. To gain insight into the nature of these phases, we developed a workflow for the characterization of PGNs using modern scanning electron microscopy (SEM) and energy-dispersive X-ray spectrometry at a low accelerating voltage of 6 kV. Automated feature analysis-a combination of morphological image analysis and elemental analysis with stage control-allowed us to detect PGNs down to 200 nm over a relatively large analysis area of 53 mm2 with a conventional silicon drift detector (SDD). Hyperspectral imaging with a high-sensitivity, annular SDD can be performed at low beam current (∼100 pA) which improves the SEM image resolution and minimizes hydrocarbon contamination. The severe overlapping peaks of the platinum group element L and M line families at 2-3 keV and the Fe and Ni L line families at <1 keV can be resolved by peak deconvolution. Quantitative elemental analysis can be performed at a spatial resolution of <80 nm; however, the results are affected by background subtraction errors for the Fe L line family. Furthermore, the inaccuracy of the matrix correction coefficients may influence standards-based quantification with pure element reference samples.